• Gas Sensors with Unique Performance

Gas sensors

Gas Sensors with Unique Performance

Some electrochemical sensors have to be operated with a bias voltage, which has the distinct disadvantage that the
baseline increases substantially with temperatures above 25°C. The HCl-sensors from Membrapor (Switzerland)
measure hydrogen chloride without biasing and show, additionally to the good baseline stability, no cross-interference
to SO2 or NO2. Due to recent efforts, the response time could be reduced down to t90 Also the NH3-sensors do not need a bias-voltage and show an exceptionally low response time. At the Sensor + Test 2007, it was shown
that the NH3/MR-100 was able to detect 1 ppm ammonia, which makes it perfectly suitable for fast leak detection with portable instruments.
An even lower detection limit is required by phosphine, which is a hazardous gas with a very low TLV-value of 0.1 ppm. In order to monitor
this gas, a highly sensitive sensor is needed. With an output signal of 4’000 nA/ppm and t90 choice for this application. It is a major task to achieve a high selectivity, but also to keep the range in cross-sensitivity to an interference gas as low as possible. A new process technology has lead to hydrogen sensors which fulfil these tasks without exception. The H2/CA-1000 has a CO-interference between 0.1 % and 2.0%, which makes this H2-sensor unique, as are many other sensors from Membrapor.

Digital Edition

IET 35.2 March

April 2025

Air Monitoring - Probe Sampling in Hazardous Areas Under Extreme Conditions - New, Game-Changing Sensor for Methane Emissions - Blue Sky Thinking: a 50-year Retrospective on Technological Prog...

View all digital editions

Events

Analytica Anacon India & IndiaLabExpo

Apr 23 2025 Mumbai, India

Analitika Expo 2024

Apr 23 2025 Moscow, Russia

Solar & Energy Storage Summit 2025

Apr 23 2025 Denver, CO, USA

Canadian Hydrogen Convention

Apr 29 2025 Edmonton, AB, Canada

Sensor + Test 2025

May 06 2025 Nuremberg, Germany

View all events