• DMA-80 Direct Mercury Analyser

Environmental Laboratory

DMA-80 Direct Mercury Analyser

Dec 02 2014

The DMA-80 can analyse solid, liquid and gas samples with equal precision over a wide dynamic range. Analysis takes only five minutes per sample and requires no sample preparation so the need for any acid digestion or wet chemistry sample treatment is eliminated. This means ease of use, low running costs and no need for hazardous chemicals to purchase, handle and dispose. Moreover, the typical bottleneck in the analytical laboratory is eliminated and also the costs usually associated with traditional mercury techniques, such as CV-AAS, ICP-AES or ICP-MS.

The DMA-80 cost of analysis is minimised thanks to the speed of analysis, catalyst and amalgamator long lifetime, sample boats durability and the possibility of using air as combustion and carrier gas.

The DMA-80 uses the principle of thermal decomposition, mercury amalgamation and atomic absorption detection. It is extremely easy to use: just weigh your sample, load it onto the built-in auto-sampler and press ‘start’.

Combining an innovative mercury measuring system with a unique optical path spectrophotometer, the DMA-80 achieves a detection limit as low as 0.001 nanograms of mercury and is capable of measuring up to a concentration of 300 mg/kg (300 ppm).

All the DMA-80 components, such as the catalytic furnace, amalgamator and spectrophotometer, are easily accessible for routine cleaning and maintenance.

The DMA-80 can be operated through a touch screen terminal or by using a standard PC. The EasyCONTROL software controls all the system functions, and provides valuable feedback on the instrument’s performances.

The DMA-80 has been used to develop the US EPA method 7473 (Mercury in solids and solutions by thermal decomposition, amalgamation, and atomic absorption spectrophotometry).

It is furthermore compliant with ASTM method D-6722-01 (Total mercury in coal and coal combustion residues) and ASTM method D-7623-10 (Total mercury in crude oil).

For the past fifteen years, the Milestone DMA-80 direct mercury analyser has offered labs around the world unmatched direct mercury analysis productivity, accuracy, safety and performance.

DMA-80 BENEFITS

Productive

• Typical sample analysis time is five minutes, with no need for acid digestion or other wet chemistry sample treatment prior to analysis.

Sensitive

• An innovative mercury measuring system with a unique optical path spectrophotometer offers extremely low detection limits.

Efficient

• Both solid and aqueous samples can be analysed with equal efficiency. A single method can be used to process organic and inorganic samples.

Convenient

• A built-in auto-sampler and easy push-button operation is appropriate for both laboratory and field settings.

Cost-Effective

• More than 70% in cost savings as compared to traditional mercury analysis techniques.

Flexible

• Easily upgraded from the basic dual-cell to tri-cell or wide-range configurations as needed.
• Add-on module available for gas analysis.


Digital Edition

IET 34.2 March 2024

April 2024

Gas Detection - Biogas batch fermentation system for laboratory use with automatic gas analysis in real time Water/Wastewater - Upcycling sensors for sustainable nature management - Prist...

View all digital editions

Events

Hannover Messe

Apr 22 2024 Hannover, Germany

FORUMESURE

Apr 22 2024 Marrakech, Morroco

Asia Water 2024

Apr 23 2024 Kuala Lumpur, Malaysia

Korea Lab 2024

Apr 23 2024 Kintex, South Korea

Canadian Hydrogen Convention

Apr 23 2024 Edmonton, AB, Canada

View all events