COD Set-Up COD VARIO

Water testing

COD Set-Up COD VARIO

14 Sep, 2010

Published over 15 years ago. See the latest and most current information on Water testing.

Waste water parameter COD

The chemical oxygen demand, ST-COD value (ST = small scale sealed tube), of a sample is determined by a dichromate method. The result can be considered as an indicator of the theoretical oxygen demand, i.e. the amount of oxygen required for the total chemical oxidation of the organic matter present in the water.

Setup CheckitDirect COD VARIO

The Lovibond® COD VARIO Set-Up allows precise water testing with minimized effort. The ST-COD concentration in mg/l O2 is measured by photometric detection.

After adding 2 ml of the sample to a Lovibond® COD VARIO tube test (LR, MR according to ISO 15705:2002), the tube is heated in the reactor for 2 hours at 150 °C, cooled down and then analysed in CheckitDirect COD VARIO photometer.

The COD Set-Up is a "ready to use" unit with a CheckitDirect COD VARIO photometer, 25 tube tests for LR and MR (each), the RD 125 thermoreactor for sample digestion and a tube stand.

Highlights

• ST-COD sealed tubes ready for use

• Suppression of chloride interference up to 1000 mg/l

• 3 ranges:

Low range:

0 - 150 mg/l, meets ISO 15705:2002

Middle range:

0 - 1500 mg/l, meets ISO 15705:2002

High range:

0 - 15000 mg/l

For further information : www.tintometer.com

Latest News

IET Guide 2026

Explore our Digital Edition

Discover the latest news and research

Digital edition

Explore Our Other Sites

Labmate Online
Mosquitoes bite with clear preferences when it comes to human scent choices
Explore more Arrow
Pollution Solutions Online
Leading UK biogas operator places first orders for new FlowSep technology
Explore more Arrow
Petro Online
Vietnam’s E10 rollout puts fuel testing capacity under closer scrutiny
Explore more Arrow
Chromatography Today
Unlock high-resolution analysis of therapeutic oligonucleotides
Explore more Arrow