Measurement and Testing
Compliant with ASTM D5453 Standard; the XPLORER-S
Apr 19 2017
International regulations regarding the maximum amount of Sulfur in fuels are getting more stringent: multi-annual programs prescribe a maximum concentration of 10 mg/kg. Best repeatability and lowest uncertainty in measuring such low sulfur concentrations is becoming a necessity.
Superior Repeatability & Accuracy
Compared to other analytical techniques, Total Sulfur measurement by UV-Fluorescence provides superior repeatability and the highest accuracy. The lowest possible uncertainty of the actual concentration provides better insight over the Sulfur content during analysis. It ultimately leads to accurate results of Sulfur concentrations to match the maximum allowed 10 mg/kg.
International methods
Current and to be implemented multi-annual programs for environmental pollution reduction, like EPA/TIER 3, CHINA 5 and EURO 5, prescribe a maximum allowable amount of 10 mg/kg Sulfur in automotive fuels. An Elemental Combustion Analyser is required to accurately measure Total Sulfur at this level as described in the international method ASTM D5453 which is comparable to ISO 20846 and SH/T 0689.
The best solution: XPLORER-S
The XPLORER-S elemental combustion analyser has been used to perform the analysis of Total Sulfur in Light Hydrocarbons according ASTM D5453. Various samples were introduced by the ARCHIE liquids auto sampler into the XPLORER-S. Accuracy, Precision and Limit of Detection tests have been executed successfully on two separate locations. One location is stationed in the Netherlands at TE Instruments and the other location in the USA at HORIBA. The results prove that the XPLORER-S succeeds with ease and is fully compliant with ASTM D5453:
PRECISION TEST – 9 mg/kg
Location 1 | Location 2 | |
CRM Standard | 9 mg/kg | 9 mg/kg |
Days | 20 | 20 |
Replicates/day | 5 | 5 |
Mean | 9.0 mg/kg | 8.7 mg/kg |
Std dev | 0.12 | 0.34 |
The Std dev is well below the allowable 0.5 mg/kg at both locations. The XPLORER-S passes the Precision Test Requirements.
ASTM D5453 Formula:
The maximum Std dev for 9 mg/kg Sulfur is 0.5 mg/kg (Std dev ≤ 1.5 * (r / 2.77)).
At the level of 9 mg/kg, r=0.93 based on the formula r=0.1788 X0.75.
ACCURACY TEST – 10 mg/kg Sulfur
Location 1 | Location 2 | |
CRM Standard | 10 mg/kg | 10 mg/kg |
Days | 10 | 10 |
Replicates/day | 3 | 3 |
Mean | 9.9 mg/kg | 9.7 mg/kg |
Std dev | 0.035 | 0.088 |
RSD (%) | 0.35 | 0.90 |
Δ | 0.1 | 0.2 |
The absolute difference (Δ) from the mean does not exceed 0.41 mg/kg. The XPLORER-S passes the Accuracy Test Requirements.
ASTM D5453 Formula:
The maximum allowable absolute difference (Δ) from the mean is calculated with the formula Max Std dev*.75.
Contact us at to see our Limit of Detection Test and other Data Reports.
Digital Edition
PIN 25.1 Feb/March
March 2024
In This Edition Safety - The technology behind the ION Science Tiger XT - Safety with ammonia and LOHCs as hydrogen carriers Analytical Instrumentation - Discussion on new tribology te...
View all digital editions
Events
Apr 03 2024 Tashkent, Uzbekistan
Salon Analyse Industrielle & Instrumentation
Apr 03 2024 Paris, France
Apr 03 2024 Atyrau, Kazakhstan
Apr 09 2024 Munich, Germany
Apr 10 2024 Helsinki, Finland