• LC-MS/MS Analysis of PFAS Extractables in Polyethersulfone (PES) Syringe Filters Using EPA 537.1
  • Figure 1. Schematic outline for testing Millex® syringe filters for PFAS contamination
  • Table 1 Detection of PFAS after filtration with nonsterile Millex® filters with PES membranes using LC/MS/MS according to EPA 537.1

PFAS Analysis

LC-MS/MS Analysis of PFAS Extractables in Polyethersulfone (PES) Syringe Filters Using EPA 537.1

May 31 2022

A key consideration for any PFAS method is to avoid contamination that can impact the accuracy of data, including those coming from sample preparation techniques such as filtration. Currently, most of the analytical methods are for “clean” matrices, such as drinking water, and often do not require filtration as a part of sample preparation. However, methods such as SW-846 Method 8327, ASTM D7968, ASTM D797 and ISO 21675 involve matrices that could have a higher degree of particulates, such as wastewater. Particulates in solution must be removed prior to LC/MS/MS, as they can be detrimental to sample analysis, column longevity and overall instrument function. These methods identify the need for filtration using membranes in a syringe filter format.

In this application note, EPA Method 537.1 was used to demonstrate that the Millex® syringe filters with PES (polyethersulfone) Millipore Express® membranes did not give any detectable levels of PFAS contamination. Figure 1 is the schematic of the experimental procedure.

No PFAS contaminants were detected even with the very low reporting limits (RL) of the method (Table 1). These results suggest that nonsterile Millex® syringe filters with PES membranes are reliable and appropriate to utilize in the filtration of samples for the analysis of PFAS compounds in environmental matrices that require filtration prior to further clean-up, by solid phase extraction for example, and/or LC-MS/MS analysis.

Follow this link to read complete application note.


Digital Edition

AET 28.3 September 2024

September 2024

Business News - ENVEA announces acquisition of APAQ Group - SICK and Endress+Hauser sign strategic partnership - Efforts to curb gas flaring intensify amid environmental concerns Air Monito...

View all digital editions

Events

INDOWATER 2024

Sep 18 2024 Jakarta, Indonesia

ICIF China 2024

Sep 19 2024 Shanghai, China

SETAC Asia-Pacific 2024

Sep 21 2024 Tianjin, China

Meteorological Technology World Expo 2024

Sep 24 2024 Vienna, Austria

View all events