• HY-OPTIMA™ 720AS-GC Hydrogen Analyzing System - Product Highlight
    HY-OPTIMA™ 720AS-GC Hydrogen Analyzing System

Gas Detection

HY-OPTIMA™ 720AS-GC Hydrogen Analyzing System - Product Highlight

Feb 26 2015

H2scan’s HY-OPTIMA 720AS-GC Hydrogen Analyzing System is designed to detect, measure and respond to hydrogen leaks in gas chromatography (GC) systems. The hydrogen specific solid-state sensing element is designed for long life, reliable operation, and for retrofit on all GC makes and models. The model 720AS-GC detects and measures hydrogen and provides alarm indications at a user selectable level. If hydrogen reaches the user selected alarm level, the alarm activates a light tower with a flashing red light, a buzzer and closes a valve to shut off the supply of hydrogen to the GC system. H2scan’s solution benefits customers and secures their safety with a quick response time, auto purge gas switch and visual & audible alarms.

Performance

Hydrogen Sensitivity Range:
0.5% to 5% hydrogen at 1 ATM on a volume basis in a single or multi-component gas background
Typical Response Time: T90 less than 60 sec
Recommended Verifiction Interval: 90 days
Product Life Expectancy: 10 years
Accuracy: ± (0.03 x indication + 0.2) percent hydrogen by volume

Instrument Display   

System Power On/Off LED
Alarm Strobe Light
Pump power On/OFF LED
Solenoid valve LED

Interface Options   

Power Suppley: 110 VAC
Serial Communications: RS-232, RS-422, USB (optional) 

Operating Conditions

Operating Humidity: 0% to 95% RH
Flow Rate: Up to 50 slpm
Operating Temperature Range: 0ºC to 40ºC
Storage Temperature Range: 0ºC to 50ºC
Calibration Background Gas: Air or Nitrogen

Alarm    

Set Point: User adjustable set point (0.5%-4%)
Response: Audible and visual output
Auto Hydrogen shut off
Inert Gas (Nitrogen, Argon) purge

Dimensions    

Length: 13 in.
Width: 15.5 in.
Depth: 3 in.

Weight

1.5 kg (~3lb.)

Product webpage


Digital Edition

IET 34.2 March 2024

April 2024

Gas Detection - Biogas batch fermentation system for laboratory use with automatic gas analysis in real time Water/Wastewater - Upcycling sensors for sustainable nature management - Prist...

View all digital editions

Events

Ozwater'23

Apr 30 2024 Melbourne, Australia

The Safety & Health Event

Apr 30 2024 Birmingham, UK

ENVEX 2024

May 03 2024 Seoul, South Korea

SETAC Europe

May 05 2024 Seville, Spain

CleanPower 2024

May 06 2024 Minneapolis, MN, USA

View all events