Auto GC Evaluation with Embedded Calibration Using Permeation Tubes

Jul 25 2016

Ambient air is polluted by many VOCs coming from Petrochemical Industry. PAMS*, Oxygenated or Halogenated VOCs can be found in many ranges of concentration between background and site fence line. These compounds are precursors of ozone and Industrial bodies need to identify them to be in compliance with the authorized limits and to improve their process.

Actually, more than 88 compounds are mentioned in US EPA list considering PAMS 56*, TO 14 or TO 15** methods. The use of calibration device is a crucial point to validate data and could be difficult when a certain amount of compounds are present.

For VOCs monitoring in ambient air, Chromatotec (France) has developed a system capable of measuring a large number of VOCs from ppt to ppm, on line and in field conditions. The device is required to identify automatically potential coeluted compounds by MS technology adapted to industrial context. The originality of the project consists in coupling two different FID GCs to a Quadripole MS allowed by an elaborated multiplexer system: one TDGC for light compounds and one TDGC for heavy compounds with specific trapping conditions and variable sampling volumes. This solution integrates internal permeation tube for automatic data validation. One using butane for C2-C6 molecule range, the other using Benzene for C6-C12.

One of the advantage of the solution is enabling to provide accurate reference points for the verification of an analysis using internal permeation tube. No needs of specific gas mixture. This solution is interesting since static mixtures in gas cylinders could be unreliable and inaccurate at the levels (ppm to ppb) that analytical technology now demands. In gas cylinders containing low level mixtures, the trace components often adsorb on to, or react with, the cylinder walls causing the standard to be inaccurate and unreliable. Also, it may be impossible to get a matrix gas that is completely free of trace component, thus making it impossible to get a true reference point with a cylinder mixture. The use of permeation tubes instead of gas cylinders offers a good alternative.

As example of use, measurement campaign on a petrochemical new site shows concentration of about 100 compounds at different steps of the commissioning. Variations during specific hours are detected. Potential coeluted compounds like terpenes or organochlorinated are identified and monitored at ppt level.

The ability of coupling in continuous two different GCs to a unique MS and the automatic identification is a new advance in industrial GC-MS technology. The big advantage is the possibility to play with two different trapping and thermo-desorption techniques linked to one MS. This fully automatic system allows non-specialist operators to access to expertise level results.

The data are validated in automatic mode thanks to combined permeation tube using butane as reference for light molecule (C2C6) and benzene for more heavy compounds (C6C12).

Actually this solution is used in autoGC evaluation driven by US EPA. Chromatotec technologies are recognised as performant in this evaluation test integrating autoGC FID with permeation tubes.

* Photochemical Assessment Monitoring Stations (PAMS)


Digital Edition

IET 34.2 March 2024

April 2024

Gas Detection - Biogas batch fermentation system for laboratory use with automatic gas analysis in real time Water/Wastewater - Upcycling sensors for sustainable nature management - Prist...

View all digital editions

Events

IFAT Munich

May 13 2024 Munich, Germany

REGATEC 2024

May 15 2024 Lund, Sweden

Disasters Expo Europe

May 15 2024 Frankurt-am-Main, Germany

AIHA Connect 2024

May 20 2024 Columbus, OH, USA

Water Expo Nigeria 2024

May 21 2024 Lagos, Nigeria

View all events